Portable Ion Chromatography CHR1A-035I
Chromatography is a technique that enables the separation, identification, and purification of the components of a mixture for qualitative and quantitative analysis. Our extensive range offers variety of products like Gas, Ion and Portable Ion chromatography products to meet all separation needs, including improved resolution, enhanced sensitivity, faster analysis and consistent performance.
Used in Food Testing, Chemical Industry, Beverage Testing, Drug testing, Forensic Science, Pharmaceutical, Molecular Biology, Medical, Research, Laboratory
Also known as Laboratory Chromatography
- Powerful data processing system:
- Iconic display, customizable interface, integration of instrument control, data analysis and processing, data sharing module for on-site and remote data sharing through 4G network.
- Quick chromatographic columns for 5-min rapid detection:
- Original quick chromatographic columns for on-site quick detection of anions and cations.
Features
- Powerful data processing system:
- Iconic display, customizable interface, integration of instrument control, data analysis and processing, data sharing module for on-site and remote data sharing through 4G network.
- Quick chromatographic columns for 5-min rapid detection:
- Original quick chromatographic columns for on-site quick detection of anions and cations.
- Intelligent flow path cleaning makes easier cleaning:
- The flow path is designed with a switching valve for free switch of eluent bottles and pure water bottles.
- WI-FI communication, real-time operation:
- Being equipped with a tablet/laptop makes real-time operation more flexibly and conveniently.
- Upgrade-supported dual detectors (Conductivity Detector and ampere detector) to meet the needs of different industries.
Specifications
Ion Chromatographic Pump: | |
Maximum Pressure | 35 Mpa (PEEK) |
Type | High-pressure and low-pulse two-piston tandem advection pump |
Flow Range | 0.001 ~ 9.999 mL/min |
Flow Accuracy | ±0.5% |
Flow Repeatability | RSD≤0.1% |
Flow Stability | (0.2-0.5) mL/min ≤ 3%; (0.5-1.0) mL/min ≤ 2%; > 1.0 mL/min ≤ 2% |
Numerical-control and Electromagnetic Sample Injector: | |
Maximum Pressure | 35 Mpa |
Control Mode | By Stepper motor |
Power Supply | 24 V (DC) |
Column Heater: | |
Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
Allowable Deviation of Column Heater’s Temperature | ± 1°C |
Temperature Stability | ≤ 0.5°C/h |
Conduction Detection System: | |
Type | Temperature control and bipolar conductivity detector |
Cell Volume | ≤0.8μL |
Detection Mode | Bipolar conductivity detection |
Detection Range | 0~45000 μS/cm (adjustable) |
Detection Resolution | ≤0.0020nS/cm |
Output Voltage | -6000~+6000 mv (adjustable) |
Baseline Noise | ≤0.5% FS |
Baseline Drift | ≤ 20% FS/30 min |
Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
Controlling Temperature Accuracy | ±0.01°C |
Maximum Pressure | 10.0 Mpa |
Instrument Linearity | ≥0.999 |
Quantitative Repeatability | ≤0.5% |
Qualitative Repeatability | ≤2% |
Minimum Detectable Concentration | Cl- ≤ 0.005 ug/mL; Li+ ≤ 0.001 ug/mL |
Flow System: | |
Six-way Valve | PEEK material, pressure 5000 psi; Independent automatic collecting and flow function. |
Panel Computer: | |
Display Screen | 12.3 inch |
Internal Memory | 2 G |
Weight | 786 g |
Maximum Pressure | 20 Mpa |
Minimum Pressure | 5 Mpa |
Suppressor: | |
Maximum Pressure | 6.0 Mpa |
Dead Volume | <30 μL |
Other Specifications: | |
Dimension (LxWxH) | 330x220x310 mm |
Net Weight | 8 kg |
Gross Weight | 11 kg |
Battery Capacity | 5000 mAh |
Power | 150 W |